In MALDI, for laser fluences below the saturation point the ion yield per shot follows a cubic dependence on the irradiated area, leading to a conclusion that smaller spots produce overall less ions and therefore are less viable. However, Qiao et al. showed that by decreasing the laser spot size it is possible to raise the saturation point, and thus increase the ion yield per unit area, also known as sensitivity. Here we explore laser spots below 10 micrometer diameter to determine whether they offer any practical advantage. We show that sensitivity is greater for a flat-top 3--4 micrometer spot than for a 10 micrometer spot. The sensitivity is greater for a Gaussian-like 3--5 micrometer spot than for flat-top 5--25 micrometer spots. We also report for the first time sensitivity versus theoretical fluence profile for a Gaussian-like beam focu.
机译:在MALDI中,对于低于饱和点的激光注量,每次发射的离子产率与辐照面积呈三次立方关系,从而得出结论,较小的斑点产生的总离子较少,因此不太可行。然而,乔等。结果表明,通过减小激光光斑的尺寸,可以提高饱和点,从而提高单位面积的离子产量,也称为灵敏度。在这里,我们探索直径小于10微米的激光点,以确定它们是否具有任何实际优势。我们显示,对于平顶3--4微米点,灵敏度要比10微米点大。高斯型3--5微米点的灵敏度比平顶5--25微米点的灵敏度更高。我们还首次报告了类似高斯光束的灵敏度与理论注量曲线的关系。