A Brillouin gain spectrometer and its evaluation by the study of some liquids.

机译:布里渊增益光谱仪及其对某些液体的研究评估。

获取原文
获取原文并翻译 | 示例

摘要

The construction and testing of a Brillouin gain spectrometer is discussed. The spectrometer employs a high power (150 mW) tunable dye laser and a low power (1.7 mW) stabilized HeNe laser in a pump-probe arrangement. Because of the small signal gain (1 {dollar}times{dollar} 10{dollar}sp{lcub}-5{rcub}{dollar}), several noise reduction schemes were used to enhance the signal to noise ratio. The spectrometer was evaluated by studying the strongly scattering liquid CS{dollar}sb2{dollar} and the weakly scattering liquid CCl{dollar}sb4{dollar}. Brillouin shifts and linewidths were accurately determined for both liquids. The high resolution of this instrument permitted measurement of the small asymmetry in the measured lineshapes. In CCl{dollar}sb4{dollar} we were able to observe the so-called Mountain mode and to confirm that CCl{dollar}sb4{dollar} can be modeled as a liquid exhibiting a single relaxation time.
机译:讨论了布里渊增益光谱仪的构造和测试。该光谱仪采用高功率(150 mW)可调染料激光器和低功率(1.7 mW)稳定的HeNe激光器,泵浦探头布置。由于信号增益小(1倍10美元10 splclc-5 rcub美元),因此使用了几种降噪方案来增强信噪比。通过研究强散射液体CS {dol} sb2 {dol}和弱散射液体CCl {dol} sb4 {dol}来评估光谱仪。两种液体均准确确定了布里渊位移和线宽。该仪器的高分辨率允许测量被测线形中的小的不对称性。在CCl {dollar} sb4 {dollar}中,我们能够观察到所谓的“山峰”模式,并确认CCl {dollar} sb4 {dollar}可以建模为表现出单个弛豫时间的液体。