Heat capacity variations in thermoluminescent dosimeter (TLD) populations have been experimentally quantified for the first time using an infrared detector. Consistent heat is applied to each sample from a TLD reader, and infrared transmission is characterized by an indium antimonide (InSb) photodiode to determine the heat capacity. Using this method, variations as high as 30% were observed among TLDs manufactured over a period of about ten years. Nearly identical trends were observed among different phosphor types (
机译:热发光剂量计(TLD)群体中的热容变化首次使用红外检测器进行了实验量化。从TLD读取器向每个样品施加一致的热量,红外传输的特征是铟锑化(InSb)光电二极管确定热容。使用这种方法,在大约十年的时间内制造的TLD之间观察到高达30%的差异。在不同类型的荧光粉之间观察到几乎相同的趋势(