首页> 外文学位 >Miniature cylindrical ion trap mass spectrometry: A study of design, optimization and functionality leading to tandem mass spectrometry.
【24h】

Miniature cylindrical ion trap mass spectrometry: A study of design, optimization and functionality leading to tandem mass spectrometry.

机译:微型圆柱离子阱质谱:导致串联质谱的设计,优化和功能的研究。

获取原文
获取原文并翻译 | 示例

摘要

This dissertation reports on the preliminary experiments showing proof of principle, the design and optimization of a miniature cylindrical ion trap operated at the nominal frequency of 1.048 MHz using software and hardware configurations designed to operate at that frequency. Consequently, mass calibrations, as well as, specialized ion manipulations were performed and reported. Results show a verified mass range of 186 m/z using hexafluorobenzene and 254 m/z using bromoform. Selective isolation of ions was successfully performed by increasing the amplitude of the rf used for storing ions. The resonance frequency of ions was determined for the first time in a miniature cylindrical ion trap. These resonance frequencies were then used for tandem mass spectrometry experiments performed on acetophenone. CID voltage and delay times were optimized for MS/MS experiments on the 120 m/z ion of acetophenone.
机译:本文对初步实验进行了报道,这些实验表明了使用设计在该频率下工作的软件和硬件配置,对在标称频率1.048 MHz下工作的微型圆柱形离子阱进行了原理验证,设计和优化。因此,进行了质量校准以及专门的离子操作并进行了报告。结果表明,使用六氟苯的确证质量范围为186 m / z,使用溴仿的质量范围为254 m / z。通过增加用于存储离子的rf的幅度成功地完成了离子的选择性隔离。在微型圆柱形离子阱中首次确定离子的共振频率。然后将这些共振频率用于对苯乙酮进行的串联质谱实验。针对在苯乙酮的120 m / z离子上进行MS / MS实验,对CID电压和延迟时间进行了优化。

著录项

  • 作者

    Meaker, Thomas Frederic.;

  • 作者单位

    Mississippi State University.;

  • 授予单位 Mississippi State University.;
  • 学科 Chemistry Analytical.
  • 学位 Ph.D.
  • 年度 2002
  • 页码 82 p.
  • 总页数 82
  • 原文格式 PDF
  • 正文语种 eng