To achieve real-time image-guided radiation therapy, the integration of a linear accelerator (linac) with a magnetic resonance (MR) imager has been proposed. This thesis presents work that investigated performance and magnetic shielding of a linac in the presence of parallel magnetic flux densities. This work used computer simulations and numerical techniques such as finite element and particle tracking algorithms. It showed that a linac can operate with only a 17 +/- 1 % target current lost and maintain its treatment beam's symmetry in the presence of 0.011 T parallel magnetic flux densities. Furthermore, the target current lost is the result of the altered electron gun optics when parallel magnetic flux densities are present. Minimal magnetic shielding (such as a 5-mm-thick, 146.5-mm-long passive shield or a pair of active shield coils with 625 and 430 A-turns) around the electron gun and waveguide was demonstrated to recover this lost target current.
机译:为了实现实时的图像引导放射治疗,已经提出了将线性加速器(直线加速器)与磁共振(MR)成像器集成在一起。本论文提出了在平行磁通密度存在的情况下研究直线加速器的性能和磁屏蔽的工作。这项工作使用了计算机模拟和数值技术,例如有限元和粒子跟踪算法。结果表明,直线加速器只能在目标电流损失为17 +/- 1%的情况下运行,并在存在0.011 T平行磁通密度的情况下保持其治疗束的对称性。此外,目标电流损耗是当存在平行磁通密度时电子枪光学系统发生变化的结果。演示了围绕电子枪和波导的最小磁屏蔽(例如5毫米厚,146.5毫米长的无源屏蔽或一对具有625和430 A匝的有源屏蔽线圈)可以恢复这种损失的目标电流。

著录项

  • 作者

    Santos, Dan Michael Uson.;

  • 作者单位

    University of Alberta (Canada).;

  • 授予单位 University of Alberta (Canada).;
  • 学科 Biophysics.
  • 学位 M.S.
  • 年度 2012
  • 页码 140 p.
  • 总页数 140
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类 老年病学;
  • 关键词

  • 入库时间 2022-08-17 11:42:49

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号