An evaluation of NCRP report 151--radiation shielding design for radiotherapy facilities, and a feasibility study for 6 MV open-door treatments in an existing high-energy radiation therapy bunker.

机译:对NCRP报告151的评估-放射治疗设施的辐射屏蔽设计,以及在现有高能放射治疗掩体中进行6 MV开门治疗的可行性研究。

获取原文
获取原文并翻译 | 示例

摘要

This thesis describes a study of shielding design techniques used for radiation therapy facilities that employ megavoltage linear accelerators. Specifically, an evaluation of the shielding design formalism described in NCRP report 151 was undertaken and a feasibility study for open-door 6 MV radiation therapy treatments in existing 6 MV, 18 MV treatment rooms at the Montreal General Hospital (MGH) was conducted. To evaluate the shielding design formalism of NCRP 151, barrier-attenuated equivalent doses were measured for several of the treatment rooms at the MGH and compared with expectations from NCRP 151 calculations. It was found that, while the insight and recommendations of NCRP 151 are very valuable, its dose predictions are not always correct. As such, the NCRP 151 methodology is best used in conjunction with physical measurements. The feasibility study for 6 MV open-door treatments made use of the NCRP 151 formalism, together with physical measurements for realistic 6 MV workloads. The results suggest that, dosimetrically, 6 MV open door treatments are feasible. A conservative estimate for the increased dose at the door arising from such treatments is 0.1 mSv, with a 1/8 occupancy factor, as recommended in NCRP 151, included.
机译:本论文描述了用于采用兆电压线性加速器的放射治疗设备的屏蔽设计技术的研究。具体而言,对NCRP报告151中描述的屏蔽设计形式进行了评估,并在蒙特利尔总医院(MGH)的现有6 MV,18 MV治疗室中对敞开式6 MV放射疗法进行了可行性研究。为了评估NCRP 151的屏蔽设计形式,对MGH的几个治疗室测量了屏障衰减等效剂量,并将其与NCRP 151计算的预期值进行了比较。已发现,尽管NCRP 151的见识和建议非常有价值,但其剂量预测并不总是正确的。因此,最好将NCRP 151方法与物理测量结合使用。使用NCRP 151形式进行6 MV开门处理的可行性研究,以及针对实际6 MV工作量的物理测量。结果表明,从剂量学上讲,6 MV开门治疗是可行的。根据NCRP 151的建议,保守估计此类治疗在门口增加的剂量为0.1 mSv,占用率为1/8。

著录项

  • 作者

    Kildea, John.;

  • 作者单位

    McGill University (Canada).;

  • 授予单位 McGill University (Canada).;
  • 学科 Nuclear physics and radiation.;Medical imaging.;Biophysics.
  • 学位 M.Sc.
  • 年度 2010
  • 页码 122 p.
  • 总页数 122
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

  • 入库时间 2022-08-17 11:36:55

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号