Microblood analyzers for the sensitive and portable detection of clinically relevant cardiac markers.

机译:Microblood分析仪用于临床相关心脏标志物的灵敏和便携式检测。

获取原文
获取原文并翻译 | 示例

摘要

Every year over 6 million patients present at hospitals with chest pains, and it is critically important to determine if AMI is the cause as quickly as possible. Conventional ELISA based tests take at least 2 hours and must be performed at a centralized location. A MEMS ELISA test has the potential to reduce these test times considerably, while performing the test cheaply and more reliably at the point of care. A MEMS PDMS device was designed along with a small, portable reader to output the results. This device carried out assays in 15 minutes with greatly increased sensitivity over conventional ELISA. The device achieved a detection range for a flow based assay of 1 pg/mL--5 ng/mL with excellent sensitivity. This device is much more sensitive than the clinically relevant cutoff level of 5 ng/mL above which AMI is considered to have occurred in the patient. Even shorter assay times can be used for less sensitive detection.; To further improve detection sensitivity, assay times and sample volume required, an improved blocking agent was developed, and was composed of Pierce protein-freeRTM blocker, 0.05% Tween 20RTM, 0.1% n-dodecyl-beta-D-maltoside, and 20 mM HEPES buffered saline. This new blocking solution was able to reduce noise from non-specific binding by ∼10 5 fold compared to ∼102 fold for a conventional blocker, BSA. This allowed a fluorescence assay to be performed on serum samples containing cFABP in PDMS devices with no need for dilution or processing to prevent interference from serum constituents. This assay achieved a sensitivity of 0.1 ng/mL with a detection range of 0.1--1000 ng/mL in 11 minutes.; A laminar tube flow with 1st order wall reaction model was used for the cFABP microdevice assay, which could predict assay behavior for a range of conditions. The pseudo-first order reaction rate was estimated at ∼1.5x10-6 m/s, and it was determined that a 40 mum wide channel operated in the desirable reaction rate limited regime, rather than the undesirable mass transfer limited regime of conventional heterogeneous immunoassays. The model also provided reasonable estimates of assay sensitivity (0.5 pg/mL--600 pg/mL depending on the assay parameters).
机译:每年有600万以上的胸痛患者在医院就诊,因此至关重要的是尽快确定AMI是否是引起疼痛的原因。基于常规ELISA的测试至少需要2个小时,并且必须在集中位置进行。 MEMS ELISA测试具有显着减少这些测试时间的潜力,同时可以在护理点以便宜且可靠的方式执行测试。设计了MEMS PDMS设备以及小型便携式阅读器以输出结果。与常规ELISA相比,该设备可在15分钟内完成检测,灵敏度大大提高。该设备实现了基于流的测定的检测范围,灵敏度为1 pg / mL--5 ng / mL。该设备比临床上相关的5 ng / mL临界值敏感得多,在该临界值以上5 ng / mL被认为是患者体内发生的AMI。甚至更短的分析时间也可以用于灵敏度较低的检测。为了进一步提高检测灵敏度,测定时间和所需的样品量,开发了一种改良的封闭剂,该封闭剂由无Pierce蛋白的RTM阻滞剂,0.05%Tween 20RTM,0.1%正十二烷基-β-D-麦芽糖苷和20 mM组成HEPES缓冲盐水。与传统阻滞剂BSA的约102倍相比,这种新的阻滞剂能够将非特异性结合产生的噪音降低约10 5倍。这样就可以在PDMS设备中对含有cFABP的血清样品进行荧光测定,而无需进行稀释或处理即可防止血清成分的干扰。该分析在11分钟内达到0.1 ng / mL的灵敏度,检测范围为0.1--1000 ng / mL。具有一阶壁反应模型的层流管用于cFABP微型设备测定,可预测一系列条件下的测定行为。拟一级反应速率估计约为1.5x10-6 m / s,可以确定40 mm宽的通道在理想的反应速率限制机制下操作,而不是常规异质免疫分析的不良传质限制机制下操作。该模型还提供了合理的测定灵敏度估算值(0.5 pg / mL--600 pg / mL取决于测定参数)。

著录项

  • 作者

    Larsen, Robert T.;

  • 作者单位

    University of Illinois at Urbana-Champaign.;

  • 授予单位 University of Illinois at Urbana-Champaign.;
  • 学科 Engineering Chemical.
  • 学位 Ph.D.
  • 年度 2007
  • 页码 177 p.
  • 总页数 177
  • 原文格式 PDF
  • 正文语种 eng