首页> 外文会议>Visualization, Image-Guided Procedures, and Display pt.2; Progress in Biomedical Optics and Imaging; vol.6,no.21 >Automated Extraction of Subdural Electrode Grid from Post-Implant MRI Scans for Epilepsy Surgery
【24h】

Automated Extraction of Subdural Electrode Grid from Post-Implant MRI Scans for Epilepsy Surgery

机译:从植入后MRI扫描中自动提取硬膜下电极网格以进行癫痫手术

获取原文
获取原文并翻译 | 示例

摘要

This paper presents an automated algorithm for extraction of Subdural Electrode Grid (SEG) from post-implant MRI scans for epilepsy surgery. Post-implant MRI scans are corrupted by the image artifacts caused by implanted electrodes. The artifacts appear as dark spherical voids and given that the cerebrospinal fluid is also dark in T1-weigthed MRI scans, it is a difficult and time-consuming task to manually locate SEG position relative to brain structures of interest. The proposed algorithm reliably and accurately extracts SEG from post-implant MRI scan, i.e. finds its shape and position relative to brain structures of interest. The algorithm was validated against manually determined electrode locations, and the average error was 1.6mm for the three tested subjects.
机译:本文提出了一种自动算法,用于从癫痫手术的植入后MRI扫描中提取硬膜下电极网格(SEG)。植入后的MRI扫描因植入的电极引起的图像伪影而损坏。这些伪像显示为黑色球形空隙,并且在T1加权MRI扫描中脑脊液也很暗,手动定位SEG相对于感兴趣的大脑结构的位置是一项困难且耗时的任务。所提出的算法可从植入后的MRI扫描可靠而准确地提取SEG,即找到其相对于感兴趣的大脑结构的形状和位置。该算法针对手动确定的电极位置进行了验证,三个测试对象的平均误差为1.6mm。