首页> 外文会议>Surface and dermal sampling >Beryllium Measurement by Optical Fluorescence in Samples Contaminated by Strongly Fluorescent Impurities
【24h】

Beryllium Measurement by Optical Fluorescence in Samples Contaminated by Strongly Fluorescent Impurities

机译:光学荧光法检测强荧光杂质污染的样品中的铍

获取原文
获取原文并翻译 | 示例

摘要

A fluorescence method for determination of beryllium and beryllium oxide particles has been approved as a standard test method by ASTM International and the National Institute for Occupational Safety and Health. The procedure involves dissolution of samples in ammonium bifluor-ide solution and adding a small aliquot of extract to a basic hydroxybenzoqui-noline sulfonate fluorescent dye and measuring its fluorescence. This method is specific to beryllium and is not affected by the presence of other metals. However, the results from such a method may be compromised by organic fluorescent impurities which have optical characteristics similar to the beryllium-bonded hydroxybenzoquinoline sulfonate. This study demonstrates that such impurities can be effectively removed by activated charcoal without compromising the sensitivity of the method or any other test attributes.
机译:用于测定铍和氧化铍颗粒的荧光方法已被ASTM International和国家职业安全与健康研究所批准为标准测试方法。该程序包括将样品溶解在氟化氢铵溶液中,并将一小部分提取物添加到碱性羟基苯并喹啉磺酸盐荧光染料中并测量其荧光。此方法特定于铍,不受其他金属的存在的影响。但是,这种方法的结果可能会受到有机荧光杂质的损害,该有机荧光杂质的光学特性类似于与铍键合的羟基苯并喹啉磺酸盐。这项研究表明,活性炭可以有效去除此类杂质,而不会影响该方法的灵敏度或任何其他测试属性。