In this paper, we present the design, implementation, and in vivo evaluation results of a wireless, batteryless, less-invasive, implantable blood-pressure-sensing microsystem for real-time monitoring of genetically engineered mice. The microsystem merges MEMS technology and low-power CMOS integrated circuit design through a high level of system integration, together with a conventional molding-based packaging technique. The implanted microsystem, consisting of a blood-pressure-monitoring unit interfaced with a low-power ASIC, can detect the blood pressure inside a vessel and wirelessly transmit the information to a nearby receiver, while using adaptive RF powering to ensure a stable system power supply. The system employs an instrumented elastic sensing cuff made of bio-compatible silicone by using a conventional molding process.
机译:在本文中,我们介绍了用于基因工程小鼠实时监测的无线,无电池,创伤小,可植入血压传感微系统的设计,实施和体内评估结果。该微系统通过高水平的系统集成以及传统的基于模制的封装技术,将MEMS技术和低功耗CMOS集成电路设计融合在一起。植入的微系统由与低功率ASIC相连的血压监测单元组成,可以检测血管内的血压并将信息无线传输到附近的接收器,同时使用自适应RF供电以确保稳定的系统功率供应。该系统采用通过常规模塑工艺由生物相容性有机硅制成的仪器化弹性感应袖带。