The purpose of the study is to characterize the imaging performance of the recently built novel cone beam breast CT (CBBCT) scanner. This CBBCT scanner system has one x-ray source and one flat panel detector (Varian's PaxScan 4030CB) mounted on a rotating assembly. A patient table is mounted above the rotating tube/detector assembly. The table has a hole through it that allows a woman's breast to hang pendant in the imaging volume at the rotation axis. The tube/detector assembly rotates around the rotation axis and acquires multiple 2D projection images of the uncompressed breast located at the rotation axis in 10 seconds. Slip ring technology allows continuous rotation of the x-ray tube/detector assembly concentric to the opening in the table to achieve multiple circle scans. Also, it has a controlledrnvertical motion during the rotation to perform a spiral scan over 20 cm of travel. The continuous 360°rotation is designed to have speeds up to 1 rev/sec. This system was validated through a series of breast-imaging phantom studies and and patient studies. The results show that the image quality of the CBBCT scanner is excellent and all phantom masses (tissue-equivalent carcinomas) and calcifications as well as human subjects' masses, calcifications and abnormalities can be detected faithfully using the CBBCT technique with a glandular dose level less than or equal to that of a single two-view mammography exam. The results indicate that the CBBCT imaging system has much better detectability of small breast tumors compared to the conventional mammography system.
机译:该研究的目的是表征最近建造的新型锥形束胸部CT(CBBCT)扫描仪的成像性能。该CBBCT扫描仪系统具有一个X射线源和一个安装在旋转组件上的平板探测器(Varian的PaxScan 4030CB)。病床安装在旋转管/检测器组件上方。桌子上有一个孔,可让妇女的乳房在旋转轴上的成像空间中悬挂吊坠。管/检测器组件绕旋转轴旋转,并在10秒内获取位于旋转轴的未压缩乳房的多个2D投影图像。滑环技术允许X射线管/检测器组件与工作台上的开口同心连续旋转,以实现多次圆形扫描。而且,它在旋转过程中具有受控的垂直运动,以在20厘米的行程上执行螺旋扫描。 360°连续旋转的速度最高可达到1转/秒。该系统已通过一系列的胸部成像体模研究和患者研究得到验证。结果表明,CBBCT扫描仪的图像质量极佳,并且可以使用CBBCT技术以较低的腺体剂量水平忠实地检测所有幻影肿块(组织等效癌)和钙化以及人类受试者的肿块,钙化和异常。等于或等于两次乳房X线检查的单次检查。结果表明,与传统的乳腺X线摄影系统相比,CBBCT成像系统具有更好的小乳腺肿瘤可检测性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号