Multiwavelength wideband intensity-modulated optical spectrometer for near-infrared spectroscopy and imaging

机译:用于近红外光谱和成像的多波长宽带强度调制光学光谱仪

获取原文
获取原文并翻译 | 示例

摘要

Abstract: Quantitative determination of chromophore concentrations by near infrared spectroscopy (NIRS) is possible by using the differential pathlength of the detected light, measured noninvasively as the mean time delay of the impulse response function of the tissue. Experimentally this information in the time domain is also available in the frequency domain by intensity modulating the input light and sweeping the modulation frequency between zero and infinity. We describe an intensity modulated optical spectrometer which differs from previously described systems in using four different wavelengths, a wideband modulation frequency (1 MHz to 500 MHz), and the simultaneous measurement of three parameters the dc intensity, ac amplitude, and the phase shift. The measured dc intensity in conjunction with the ac phase shift data allows changes in absorption coefficient (and hence chromophore concentration) to be determined more accurately by correcting for real time path length variations. The ac phase shift in combination with the ac modulation depth theoretically allows for the absolute measurement of tissue absorption and scattering coefficient. Preliminary performance figures for the system suggest values of rms noise of 0.0006 OD, 0.0011 rad and 0.0008% for the measured attenuation, ac phase shift and modulation depth. Using a phantom of fixed geometry with known scattering and absorption properties, the ability of the system to reproduce the information content of the impulse response function for a homogeneous phantom is investigated. !16
机译:摘要:利用检测到的光的差光程长度(通过非侵入式测量作为组织的脉冲响应功能的平均时间延迟),可以通过近红外光谱(NIRS)定量测定生色团浓度。实验上,通过强度调制输入光并在零和无穷大之间扫描调制频率,可以在频域中获得时域中的此信息。我们描述了一种强度调制光学光谱仪,它与先前描述的系统的不同之处在于使用了四个不同的波长,一个宽带调制频率(1 MHz至500 MHz)以及同时测量了三个参数(直流强度,交流幅度和相移)。测得的直流强度与交流相移数据相结合,可以通过校正实时路径长度变化来更准确地确定吸收系数(进而生色团浓度)的变化。理论上,交流相移与交流调制深度相结合,可以绝对测量组织的吸收和散射系数。该系统的初步性能数据表明,对于测量到的衰减,交流相移和调制深度,均方根噪声值为0.0006 OD,0.0011 rad和0.0008%。使用具有已知散射和吸收特性的固定几何体模型,研究了系统针对均匀模型复制脉冲响应函数信息内容的能力。 !16

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号