MICRO-COULOMETRY FOR MEASURING OXYGEN CONTENT IN COPPER OXIDES

机译:微库仑法测定铜氧化物中的氧含量

获取原文
获取原文并翻译 | 示例

摘要

A micro-coulometry system was newly developed and employed for measuring the oxygen content in copper oxides using a reduced amount of material. To achieve this reduction in sample size, Ar flow rate, and coulometric current were optimized. When using 5.0 mg of YBa_2Cu_3O_(6.71), which is about 1/10 the amount used in conventional coulometry, the oxygen content was successfully determined to be 6.68 ± 0.03 by correcting the side reaction. It was found to be necessary to continue to pass Ar through the solution before coulometry more than ten minutes and during the coulometric measurement. This method will be useful for measuring the oxygen content of oxide samples smaller than 10 mg, i.e. oxide materials produced in small quantities for electronic applications.
机译:新开发了一种微库仑法系统,并使用减少的材料量来测量氧化铜中的氧含量。为了实现样品量的减少,对Ar流量和库仑电流进行了优化。当使用5.0 mg的YBa_2Cu_3O_(6.71)(约为常规电量分析法中所用量的1/10)时,通过校正副反应成功确定了氧含量为6.68±0.03。已发现在电量分析前十分钟以上和电量分析测量期间必须使Ar继续通过溶液。该方法对于测量小于10 mg的氧化物样品(即为电子应用少量生产的氧化物材料)的氧含量将很有用。