首页> 外文会议>Optical Fabrication and Testing >Application of a deflectometry method to deep aspheric ophthalmic surface testing
【24h】

Application of a deflectometry method to deep aspheric ophthalmic surface testing

机译:折光法在深非球面眼科检查中的应用

获取原文
获取原文并翻译 | 示例

摘要

Abstract: Optical surfaces control poses a specific problem inthe case of ophthalmic surfaces, particularly forprogressive addition surfaces. Usual interferometrymethods do not allow to measure easily suchmeasurements but the duration and the sophisticatedenvironment needed increase too much the testing cost.For that purpose, we have developed and patented aprogressive addition surface measurement system using adeflectometry method. THe Ronchi method we use consistsin illuminating the simple by a known wavefront and inmeasuring the slopes of the reflected rays. The mainissue is then to compute accurately the surfacetopography. To solve this problem, we use anoptimization method. Finally, the built instrumentallows us to measure automatically deep asphericsurfaces in a few minutes with an accuracy close to 0.2percent of its asphericity. It doesn't need acontrolled environment. !7
机译:摘要:对于眼科表面,特别是对于渐进式加法表面,光学表面控制提出了一个特定的问题。常规的干涉测量方法无法轻松测量此类测量,但是所需的持续时间和复杂的环境增加了太多的测试成本。为此,我们开发了使用偏转测量方法的渐进附加表面测量系统并申请了专利。我们使用的朗奇方法包括通过已知的波前照亮简单波并测量反射光线的斜率。然后,主要问题是准确计算表面形貌。为了解决这个问题,我们使用了一种优化方法。最后,这种内置仪器使我们能够在几分钟内自动测量深非球面,其精度接近其非球面度的0.2%。它不需要受控的环境。 !7