Abstract: This paper describes the use of a Hartmann-type pupil mask and CCD camera to perform wavefront and focus validation of the Hubble Space Telescope simulator before and during environmental testing of the second-generation Wide Field and Planetary Camera. The method yields a focus accuracy at F/24 of about $POM 100 micrometers even in the presence of 3.5 waves of surface spherical aberration. The method avoids the introduction of potentially imperfect auxiliary optical tooling (e.g., null correctors). !7
机译:摘要:本文介绍了在第二代广角和行星相机进行环境测试之前和期间,使用Hartmann型瞳孔掩模和CCD相机对哈勃太空望远镜模拟器进行波前和聚焦验证的过程。即使在表面波像差为3.5波的情况下,该方法在F / 24时的聚焦精度也约为100微米约$ POM。该方法避免了引入可能不完美的辅助光学工具(例如,空校正器)。 !7