In this paper, false nearest adjacent points method will be used for error compensation of heating furnace in device for standard gas with trace water preparation. In different temperature difference between environment and heating furnace, phase space of nonlinear time series for temperature will be reconstructed, which can improve the sensitivity of the input data for NN and simplify the measurement model. Then nonlinear regression will be completed by RBF NN. So the temperature measurement model of heating furnace will be obtained. Experiment proves that the measurement error in this method is reduced greatly, also the influence of environment temperature is reduced. By the contrast experiment, the good performance of the FNN-RBFNN is verified.
机译:本文在本文中,最近的相邻点法将用于痕量水制剂的标准气体中加热炉的误差补偿。在环境和加热炉之间的不同温差中,将重建温度的非线性时间序列的相位空间,这可以提高NN的输入数据的灵敏度,并简化测量模型。然后,非线性回归将由RBF NN完成。因此,将获得加热炉的温度测量模型。实验证明,该方法中的测量误差大大减少,也减少了环境温度的影响。通过对比试验,验证了FNN-RBFNN的良好性能。