The Error Analysis of Automated Biochemical Analyzer

机译:自动化生物化学分析仪的误差分析

获取原文
获取外文期刊封面目录资料

摘要

In order to find out the relationships between the four error sources, namely, locating error, optics system fluctuations, sampling error and temperature fluctuations, and the accuracy of the instrument, we analyzed each relationship of them, and found that the effect of locating error is so tiny that it can be ignored, finally established the total error model and the model of the allowed noise of the optics system of the instrument. Analysis and the models may supply a reliable theory evidence for the improvement of the instrument and the designing of similar instruments.
机译:为了了解四个错误源之间的关系,即定位误差,光学系统波动,采样误差和温度波动以及仪器的准确性,我们分析了它们的每个关系,发现定位错误的效果这太小了,可以忽略它,最后建立了总误差模型和仪器光学系统允许噪声的模型。分析和模型可以提供可靠的理论证据,了解仪器的改进和类似仪器的设计。