首页> 外文会议>International Conference on 3D Radiation Dosimetry >Dosimetric impact of a change in breathing period on VMAT stereotactic ablative body radiotherapy
【24h】

Dosimetric impact of a change in breathing period on VMAT stereotactic ablative body radiotherapy

机译:呼吸周期变化对vmat立体定向烧蚀体放射治疗的剂量影响

获取原文

摘要

The dosimetric impact of a change in breathing period during treatment was assessed for a volumetric modulated arc therapy (VMAT) stereotactic ablative radiotherapy (SABR) lung plan optimized according to our centre's planning protocol. Plan delivery was evaluated at three breathing rates ranging from 7 to 23 breaths-per-minute (BPM) against the planning anatomy (15 BPM) calculated dose. Dynamic ion chamber, EBT3 film and Fricke-xylenol orange-gelatin (FXG) gel measurements were acquired using a motion phantom with appropriate inserts for each dosimeter. The results show good agreement between measured and calculated plan dose within the internal gross tumour volume (IGTV) target.
机译:根据我们的中心的规划协议,评估治疗期间呼吸周期变化的剂量造成的呼吸周期变化的影响,用于根据我们的中心的规划方案进行优化的体积调制弧治疗(VMAT)肺计划。计划递送以三种呼吸速率评估,从7至23次呼吸每分钟(BPM),针对规划解剖学(15bpm)计算的剂量。使用具有适当插入物的运动模型获得动态离子室,EBT3膜和Fricke-Xylenol橙 - 明胶(FXG)凝胶测量。结果表明,在内部总肿瘤体积(IGTV)靶标中测量和计算的计划剂量之间的良好一致性。