首页> 外文会议>International Conference on 3D Radiation Dosimetry >Dosimetry of Strontium eye applicator: Comparison of Monte Carlo calculations and radiochromic film measurements
【24h】

Dosimetry of Strontium eye applicator: Comparison of Monte Carlo calculations and radiochromic film measurements

机译:锶眼涂抹器的剂量测定:蒙特卡罗计算和放射性致铬膜测量的比较

获取原文
获取外文期刊封面目录资料

摘要

Strontium-90 eye applicators are a beta-ray emitter with a relatively high-energy (maximum energy about 2.28 MeV and average energy about 0.9 MeV). These applicators come in different shapes and dimensions; they are used for the treatment of eye diseases. Whenever, radiation is used in treatment, dosimetry is essential. However, knowledge of the exact dose distribution is a critical decision-making to the outcome of the treatment. The main aim of our study is to simulate the dosimetry of the SIA.20 eye applicator with Monte Carlo GATE 6.1 platform and to compare the calculated results with those measured with EBT2 films. This means that GATE and EBT2 were used to quantify the surface and depths dose-rate, the relative dose profile and the dosimetric parameters in according to international recommendations. Calculated and measured results are in good agreement and they are consistent with the ICRU and NCS recommendations.
机译:锶-90眼涂器是一种β射线发射器,具有相对高能量的发射器(最大能量约为2.28meV和平均能量约为0.9 meV)。这些涂抹器具有不同的形状和尺寸;它们用于治疗眼病。每当,辐射用于治疗中,剂量测定是必不可少的。然而,对确切剂量分配的了解是对治疗结果的关键决策。我们的研究的主要目的是模拟SIA.20眼涂器的剂量测定与蒙特卡罗门6.1平台,并将计算结果与用EBT2薄膜测量的那些进行比较。这意味着根据国际建议使用栅极和EBT2来量化表面和深度剂量率,相对剂量曲线和剂量分析参数。计算和测量结果符合良好的一致性,它们与ICRU和NCS建议一致。