首页> 外文会议>Symposium on static and dynamic spinal implants >Comparison of Self-Drilling and Self-Tapping Cervical Spine Screws Using ASTM F543-07
【24h】

Comparison of Self-Drilling and Self-Tapping Cervical Spine Screws Using ASTM F543-07

机译:使用ASTM F543-07对自钻和自攻颈椎脊柱螺钉的比较

获取原文

摘要

ASTM F543-07 is used to evaluate insertion load (Annex A4), driving torque (A2), and axial pullout (A3) characteristics of self-tapping metallic medical bone screws. F543-07 was designed to evaluate screws that utilize a pre-drilled pilot hole and was not intended for self-drilling screws. The objectives of this study were to (1) adapt the F543-07 test methods to allow evaluation of both self-drilling and self-tapping screws and (2) quantify differences in performance between the two screw types. For all three tests (A4, A2, A3) appropriately sized pilot holes were pre-drilled into the surrogate test medium for all the self-tapping screw designs per ASTM F543-07. No pilot holes were created for the self-drilling screws. During the insertion load tests (A4), a constant rotation was applied to the screws under an initial, near-zero axial force state. While monitoring axial load, torque, and displacement, the axial load was gradually increased. The maximum load at which there was a marked increase in torque and displacement on the graph was indicative of the force required to engage the self-drilling or the self-tapping feature. To measure driving torque (A2), a constant rotation was applied. To maintain screw driver-screw contact, a constant-force axial pre-load was required. The axial insertion load, determined during the insertion load testing (A4), was used as the preload for the driving torque testing. Axial pullout testing (A3) measured the axial tensile force required to remove a screw from the test medium. Quantifiable differences in mean insertion load, driving torque, and pullout load were found, demonstrating that these methods are able to discern differences in performance between different screw designs. Finally, this study demonstrated it is possible to adapt the test methods outlined in ASTM F543-07 to evaluate the insertion load, driving torque, and axial pullout performance of self-drilling screws in addition to self-tapping screws.
机译:ASTM F543-07用于评估自攻金属医用骨螺钉的插入载荷(附件A4),驱动扭矩(A2)和轴向拉出(A3)特性。 F543-07旨在评估利用预钻的先导孔的螺钉,而不是用于自钻螺钉。本研究的目标是(1)调整F543-07测试方法,以允许评估自钻和自攻螺钉和(2)量化两个螺钉类型之间的性能差异。对于所有三个测试(A4,A2,A3)适当大小的导向孔被预先钻入替代测试介质,用于所有ASTM F543-07的所有自攻螺杆设计。为自钻螺钉创建了导孔。在插入负载测试(A4)期间,在初始近零轴力状态下施加恒定旋转。虽然监测轴向载荷,扭矩和位移,但轴向载荷逐渐增加。在图中有扭矩和位移的标记增加的最大负载表明接合自钻或自攻特征所需的力。为了测量驱动扭矩(A2),施加恒定旋转。为了保持螺杆驱动螺杆触点,需要恒定力的轴向预载荷。在插入负载测试(A4)期间确定的轴向插入负载用作驱动扭矩测试的预载。轴向拉出测试(A3)测量了从测试介质中除去螺杆所需的轴向拉伸力。找到了平均插入负载,驱动扭矩和拉出载荷的可量化差异,表明这些方法能够在不同螺杆设计之间的性能差异辨别出来。最后,本研究证明,除了自攻螺钉之外,还可以调整ASTM F543-07中概述的测试方法,以评估自钻螺钉的插入负载,驱动扭矩和轴向拉出性能。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号