首页> 外文会议>International Conference on Advancements in Nuclear Instrumentation, Measurement Methods and their Applications >Attractive Approach to Analyse Trace and Ultra-trace Elements on Transmutation Targets by Direct Separation of Isobaric Interferences using Collision Reaction Cell Implanted on MC-ICPMS
【24h】

Attractive Approach to Analyse Trace and Ultra-trace Elements on Transmutation Targets by Direct Separation of Isobaric Interferences using Collision Reaction Cell Implanted on MC-ICPMS

机译:通过在MC-ICPMS上植入的碰撞反应细胞直接分离因果干扰来分析嬗变目标对嬗变目标的吸引力和超微量元素

获取原文

摘要

The experiments involved in the PHENIX french nuclear reactor to obtain precise and accurate data on the total capture cross sections of the heavy isotopes and fission products require isotopic ratios measurements with uncertainty of a few per mil. These accurate isotopic ratio measurements are performed with mass spectrometer equipped with multicollector system. The implantation of collision-reaction cells on Multiple-Collector Inductively Coupled Plasma Mass Spectrometer offers the possibility to separate directly in the cell several interfering elements. We present three applications of direct separation in a collision reaction cell in the case of three transmutation targets: ~(23S)U, ~(153)Eu and ~(95)Mo. Three different reactions gas NH_3, O_2 and N_2O were used to separate respectively Pu-U, Eu-Gd-Sm and Mo-Zr interfering elements. This approach is of great interest to reduce the risk of natural contamination, the amount of analytical radioactive waste, and the time contact of analysts with radioactive samples.
机译:参与苯法核反应堆的实验,以获得重型同位素和裂变产物的总捕获横截面的精确和准确数据,需要具有每密耳的不确定性的同位素比率。这些精确的同位素比率测量是用配备有多个负块系统的质谱仪进行。植入碰撞 - 反应电池对多集电感应耦合等离子体质谱仪的植入能够直接分离在细胞几个干扰元件中。我们在三个嬗变目标的情况下展示了三种碰撞反应细胞中的直接分离:〜(23s)U,〜(153)欧盟和〜(95)Mo。三种不同的反应气体NH_3,O_2和N_2O分别分离PU-U,EU-GD-SM和MO-ZR干扰元件。这种方法很有兴趣,以降低自然污染的风险,分析器放射性废物的量,以及分析人员与放射性样品的时间接触。

著录项

相似文献

  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号