首页> 外文会议>NCSL International Workshop and Symposium >Optimization of Thermal Radiation Source for High Temperature Infrared Thermometer Calibration
【24h】

Optimization of Thermal Radiation Source for High Temperature Infrared Thermometer Calibration

机译:高温红外温度计校准的热辐射源优化

获取原文
获取外文期刊封面目录资料

摘要

Industrial level infrared thermometers are being increasingly designed to measure temperatures above 500°C. A thermal radiation source is needed to calibrate these instruments. The infrared thermometers designed to measure these temperatures generally measure with a smaller field-of-view. This means there is a possibility of using a blackbody cavity as the thermal radiation source. A previous attempt was made to use a cavity mounted inside of a furnace meant for thermocouple calibration for a thermal radiation source. However, when the cavity's emissivity was measured, it was found the emissivity was not constant for different wavelengths and was very dependent on cavity position. This paper discuses a new attempt to mount a cavity inside a thermocouple furnace with much better temperature uniformity. It discusses the measurements made to verify the emissivity of the cavity using industrial radiation thermometers. It also talks about measurements made with industrial handheld infrared thermometers and compares these measurements to measurements made using other thermal radiation sources and the associated measurement uncertainty.
机译:工业水平红外温度计越来越旨在测量500°C以上的温度。需要热辐射源来校准这些仪器。旨在测量这些温度的红外温度计通常用较小的视野测量。这意味着有可能使用黑体腔作为热辐射源。先前的尝试使用安装在炉内的空腔,意为热辐射源的热电偶校准。然而,当测量腔的发射率时,发现发射率不是不同波长的恒定,并且非常依赖于腔位置。本文消除了一种新的尝试,以便在热电偶炉内安装腔,以更好的温度均匀性。它讨论了使用工业辐射温度计验证腔的发射率的测量。它还讨论了用工业手持式红外温度计进行的测量,并将这些测量与使用其他热辐射源和相关的测量不确定性进行比较。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号