首页> 外文会议>European Conference on Biomedical Optics >Tissue Oxygenation during Exercise Measured with NIRS:A Quality Control Study
【24h】

Tissue Oxygenation during Exercise Measured with NIRS:A Quality Control Study

机译:使用NIR测量运动期间的组织氧合:质量控制研究

获取原文

摘要

We assess the data quality of calculated tissue oxygen saturation (SO_2) and haemoglobin concentrations recorded on muscle during an incremental cycling protocol in healthy volunteers. The protocol was repeated three times at the same day and a fourth time at a different day to estimate the reproducibility of the method. A novel broad-band, spatially resolved spectrometer (SRS) system was employed which allowed us to compare SRS–based oxygenation parameters with modified Lambert-Beer (MLB) data. We found that the inter-subject variation in SO_2 (standard deviation about 6 %) is considerably larger than the reproducibility (about 1.5 %) both for same day and different day tests. When changes in SO_2 during the cycling test were considered the reproducibility is better than 1 %. Time courses of SRS–based haemoglobin parameters are different from MLB–data with higher reproducibility for SRS. The magnitudes of the haemoglobin changes were found to be considerably larger for the SRS method. Furthermore, the broad band approach was tested against a four-wavelength analysis with the differences found to be negligible.
机译:在健康志愿者的增量循环协议期间,评估计算出的组织氧饱和度(SO_2)和血红蛋白浓度的数据质量。在同一天重复该方案并在不同的一天重复三次,以估计该方法的再现性。采用新型宽带空间分辨光谱仪(SRS)系统,其使我们能够将基于SRS的氧合参数与改性兰伯特 - 啤酒(MLB)数据进行比较。我们发现SO_2的互受对象变化(标准偏差约6%)比同日和不同日期测试的再现性(约1.5%)大大大。当循环试验期间SO_2的变化被认为是重现性优于1%。基于SRS的血红蛋白参数的时间课程不同于MLB数据,具有较高的SRS再现性。对于SRS方法,发现血红蛋白变化的大小相当大。此外,对四波长分析测试了宽带方法,发现差异可忽略不计。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号