摘要

This paper presents the method of intraocular pressure (IOP) determination by transpalpebral tonometry (TPT), especially for the diagnosis and the control of glaucoma. TPT can be preferred method for determining IOP in terms of minimizing the contact of the measuring device with the surface of the eye. The biomechanical basis of transpalpebral tonometry is briefly discussed. The two possible schemes of TPT technique which includes a stem with springs as measuring sensor are presented. Modeling of TPT system for parameters optimization of the measuring sensor in case the dynamic measurement method is described. The researches conducted with various combinations of stem weight and diameters are presented for IOP in the range from 10 to 30 mmHg.
机译:本文介绍了转骨膜眼压术(TPT)的眼压(IOP)测定方法,特别是对于诊断和荧光眼的控制。在最小化测量装置与眼睛的表面的触点最小化的方面,TPT可以是优选的方法。简要讨论了Transpalpebral Tononetry的生物力学基础。呈现了作为测量传感器的具有弹簧的杆的TPT技术的两种可能方案。描述了在描述动态测量方法的情况下测量传感器参数优化的TPT系统建模。用各种茎重和直径组合进行的研究在10至30mmHg的范围内呈现IOP。