Maintaining Sensitivity and Reproducibility with the JetCleanJetCleanSelfSelf--Cleaning Cleaning Ion Source for Pesticides in Food Ion Source for Pesticides in Food and FeedFeed

机译:在食品和饲料中的食品离子源中的杀虫剂中的杀虫剂的敏感性和再现性保持敏感性和再现性

获取原文

摘要

Approximately 170 various pesticides were analyzed in Organic Honey on the 7010 Series Triple Quadrupole GC/MS using standard and JetClean source in the Acquire and Clean mode, utilizing carefully introduced hydrogen flow. The JetClean control is included in the MassHunter software with easy setup and operation. The chromatographic peak shape and baseline was improved using the Jetclean source particularly for the late eluting compounds. The calibration resulted very comparable R~2 values by both source, while the MDLs obtained at 1.25 pg level resulted lower values using the JetClean source. The %RSDs were comparable at higher, 2.5pg/μl level. The results indicate that the JetClean source meets and exceed the performance delivered by the standard source. Further study is undergoing to identify how the source maintenance period is extended when JetClean is applied compared to the standard source.
机译:在7010系列Triple Quadrupole GC / MS上使用标准和清洁模式进行了大约170种各种农药,在7010系列三重四极杆GC / MS中,利用仔细引入的氢气流动。 JetClean控制器包含在Masshunter软件中,具有易于设置和操作。使用Jetclean源改善了色谱峰形状和基线,特别是对于晚洗脱化合物。校准由两个源产生非常相当的R〜2值,而使用JetClean源的1.25pg水平获得的MDL产生较低的值。 %RSD在较高,2.5pg /μl水平上相当。结果表明,JetClean源符合并超过标准源提供的性能。正在进行进一步的研究以确定如何在与标准源相比应用JetClean时源维护期的延长。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号