首页> 外文会议>NORCHIP Conference >Integrated circuit for neural recording and stimulation
【24h】

Integrated circuit for neural recording and stimulation

机译:用于神经记录和刺激的集成电路

获取原文

摘要

A 16 channel neural signal measurement and stimulation system is presented. All 16 channels can be used in a measurement setup that simultaneously both stimulates and measures neurons. The amplifying stage consists of 16 preamplifiers, a multiplexer and two post-amplifiers. The measured signal is digitized with a 10-bit SAR A/D converter. The circuit contains control logic for buffering the measured results and inserting them into a SPI-bus. The logic also controls the circuit's bias currents, high-pass corner frequency, post-amplifier gains and stimulation current strengths. Amplifying gain can be configured from 100 to 5000. The mixed-signal circuit is implemented with 0.35 μm IC process by ams AG. The circuit is estimated to consume from 900 μW to 5 mW power, depending on the SPI-bus clock speed and amplifying gain.
机译:提出了16个通道神经信号测量和刺激系统。所有16个通道都可以用于测量设置,同时刺激和测量神经元。放大阶段由16个前置放大器,多路复用器和两个后放大器组成。测量信号用10位SAR A / D转换器数字化。该电路包含控制逻辑,用于缓冲测量结果并将其插入SPI-总线。逻辑还控制电路的偏置电流,高通角频率,放大器发电机收益和刺激电流优势。放大增益可以从100到5000配置。混合信号电路通过AMS AG使用0.35μm的IC过程实现。根据SPI总线时钟速度和放大增益,估计电路估计消耗900μW至5 MW功率。