首页> 外文会议>International Symposium on Ceramics in Medicine >Clinical wear and simulator study of ceramic-ceramic THR to 20years and beyond
【24h】

Clinical wear and simulator study of ceramic-ceramic THR to 20years and beyond

机译:陶瓷陶瓷Thr到20年及以后的临床磨损和模拟器研究

获取原文

摘要

This study is the first report to grade the micro wear progress in clinical and simulator wear of ceramic on ceramic bearings to equivalent of 20years use. Implants were retrieved after 15-25 years of successful use and compare to 10 to 20 million cycle simulator studies. This report examines the ball and cup wear seen on 2 Mittelmeier retrievals in the USA with that seen on a conventional all-ceramic THR retrieved in Vienna after 24 years. In SEM study wear severity was classified into 5 grades. In retrieved implants gray coloring matter was deposited on load-bearing area. Gray stripe areas were observed on the periphery of the head. The SEM analysis showed that the main bearing area had high wear (Grade 4). The peripheral gray stripe area had sever wear (Grade 5). EDAX showed that gray color was caused by transferred CoCr particles. In the simulator study the wear progress was most sever in main load-bearing area (Grade 4). All-ceramic implants had minimal wear even after 24 years. The surface of retrieved implants had different types of wear. The contact zones showed normal wear and showed relatively mild. In the periphery of the contact zones, the stripe scars corresponded with the cup rim tracking and more sever wear. Such stripe wear was not observed in the simulator experiments. The gray wear areas were caused by the metal contamination from micromotion of stem against bone. Thus, ceramic bearings proved to be excellent after 15-25 years by simulator studies and clinical studies.
机译:本研究是第一份植入陶瓷轴承陶瓷临床和模拟器磨损的微观磨损进展的报告,相当于20年的使用。在成功使用15-25年后,植入物检出并比较10至2000万周期模拟器研究。本报告探讨了美国24年后在维也纳检索的传统全陶瓷Thr中看到的2 Mittelmeier检索的球和杯子。在SEM学习中,磨损严重程度被分为5等级。在检索到的植入物中,灰色着色物沉积在承载区域上。在头部的周边观察到灰色条纹区域。 SEM分析表明,主轴承区域具有高磨损(4级)。周边灰色条纹区域具有切割磨损(5级)。 edax表明灰色颜色是由转移的Cocr颗粒引起的。在模拟器研究中,主承载区域(4级)中的磨损进度大多数。甚至在24年后,全陶瓷植入物的磨损性最小。检索植入物的表面具有不同类型的磨损。接触区显示正常磨损并显示相对温和。在接触区域的周边中,条纹疤痕与杯子轮辋跟踪和更断的磨损相对应。在模拟器实验中未观察到这种条纹磨损。灰色磨损区域是由茎骨微调的金属污染引起的。因此,通过模拟器研究和临床研究,陶瓷轴承被证明在15-25年后优异。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号