Establishment and maintenance of resistance standard traceability is one of the main tasks of Primary Electromagnetic Laboratory (PEL) within the Faculty of Electrical Engineering and Computing of the University of Zagreb. For the purpose of accurate resistance standards comparison a digital voltmeter (DVM) based method has been designed and realized. When measuring resistance standards greater than 10 MΩ capabilities of the method are restricted due to influence of DVM's input parameters: input DVM's resistance, input offset current and input capacitance. Determination of DVM's input parameters and their contribution to the overall high-ohm resistance comparison uncertainty has been presented in this paper. The possibility of use of DVM-based method for gigaohm value resistors comparison has also been investigated.
机译:抵抗标准可追溯性的建立和维护是萨格勒布大学电气工程和计算中主要电磁实验室(PEL)的主要任务之一。为准确的电阻标准比较,设计和实现了基于数字电压表(DVM)的方法。由于DVM的输入参数的影响,测量该方法的电阻标准大于10mΩ的方法:输入DVM的电阻,输入偏移电流和输入电容。本文提出了对DVM的输入参数的确定及其对整体高欧姆电阻比较不确定性的贡献。还研究了利用基于DVM的方法的千兆数值电阻比较的可能性。