Study of non-contact type flowmeter for low flow rates using near-infrared light

机译:使用近红外光进行低流速的非接触式流量计的研究

获取原文
获取外文期刊封面目录资料

摘要

We have proposed a novel method of non-contact thermal type measurement of low flow rate using the change in the absorption spectrum of water in the near-infrared wavelength range. With a prototype flowmeter using two detectors, we have made experiments and obtained calibration curves between the estimated and true flow rates. From the experimental results, we can measure the flow rates from 0.5 to 9 ml/min with the accuracy of about 10% of the flow rate. We also made a simulation of the flowmeter, and obtained agreement between the simulation and measurement. Discussions are made to improve the measurement accuracy.
机译:我们已经提出了一种新的非接触热型测量方法,使用近红外波长范围内的水的吸收光谱的变化的变化。使用使用两个探测器的原型流量计,我们已经进行了实验并获得了估计和真正流速之间的校准曲线。从实验结果中,我们可以测量0.5至9毫升/分钟的流速,精度约为流速的10%。我们还对流量计进行了模拟,并在模拟和测量之间获得了一致性。进行讨论以提高测量精度。