首页> 外文会议>Annual International Conference of the IEEE Engineering in Medicine and Biology Society >CAMPI: computerized analysis of mammography phantom images-effects of film nonlinearity and digitizer noise
【24h】

CAMPI: computerized analysis of mammography phantom images-effects of film nonlinearity and digitizer noise

机译:Campi:乳房X线摄影幻影图像的计算机化分析 - 薄膜非线性和数字仪噪声的影响

获取原文

摘要

The purpose of this work was to extend the recently introduced CAMPI method to allow for the nonlinear sensitometric response of film. Sensitometric strips were digitized and used to calibrate the phantom images by converting the pixel value to a linearized intensity-like variable. The corresponding linearized CAMPI measures were found to have simpler (and expected) dependencies on X-ray variables as long as the optical density of the film was not too large. At higher densities (2.0) the digitizer noise becomes significant, and a correction was necessary. A simple model for the digitizer noise was proposed and tentatively verified. With this work CAMPI can now perform objective and accurate evaluations on analog and direct digital images, compensating for phantom variations, mAs differences, screen-film characteristics (if applicable), and even allowing for digitizer noise limitations.
机译:这项工作的目的是扩展最近介绍的Campi方法,以允许薄膜的非线性敏感响应。通过将像素值转换为线性化强度的变量来校准敏感度表并用于校准幻像图像。发现相应的线性化Campi措施在X射线变量上具有更简单的(和预期的)依赖性,只要薄膜的光学密度“不太大。在较高的密度(<2.0)处(<2.0),数字转换器噪声变得显着,并且需要校正。提出了一个简单的数字磁噪声模型,并暂时验证。通过这项工作,Campi现在可以对模拟和直接数字图像进行客观和准确的评估,补偿幻影变化,MAS差异,屏幕胶片特性(如果适用),甚至允许数字转换器噪声限制。