Coil electrodes for RF ablation of atrial fibrillation

机译:用于RF消融心房颤动的线圈电极

获取原文

摘要

Long and contiguous lesions may cure AF. We analyzed a multielectrode steerable catheter capable of creating such lesions. For optimal temperature control purposes, based on analytical and experimental data, one temperature sensor was placed at each end of the long coil electrodes. In vitro ablations were performed to analyze the catheter performance. At set temperatures of 70/spl deg/C, 85% of the lesions were contiguous. All lesions created at set temperatures of 80 and 90/spl deg/C were contiguous. No sudden impedance rises occurred. The lesion length increased linearly with the number of electrodes. Power to reach set temperature was larger and the impedance decreased as more electrodes were connected in the ablation circuit.
机译:长且邻接的病变可以治愈AF。我们分析了一种能够产生这种病变的多电极可转向导管。为了获得最佳温度控制目的,基于分析和实验数据,将一个温度传感器置于长线圈电极的每个端部。进行体外消融以分析导管性能。在70 / SPL DEG / C的设定温度下,85%的病变是连续的。在80和90 / SPL DEG / C的设定温度下产生的所有病变都是连续的。没有发生突然阻抗。病变长度与电极数量线性增加。达到设定温度的功率较大,并且随着更多电极在消融电路中连接的阻抗降低。