Under the line-source model, ensembles of concentric-needle motor unit action potentials (MUAPs) were simulated. Neuromuscular junction jitter was simulated by delaying the initiation of muscle fibre action potentials (MFAPs) by a normal random variate. To transform the MFAP contributions to a MUAP into symmetrical, singly-peaked, positive functions, inverse-average-current filtering was applied to the MUAPs of the ensembles, producing composite weight functions. The jitter estimate, defined as the standard deviation of composite weight function interpeak intervals over an ensemble, was correlated to the standard deviation of the random variable used to delay the initiation of the MFAPs. Since normal jitter is of the order of 25 μsec, the base sampling interval of 40 μsec was enhanced to approximately 2 μsec using divided-difference interpolation. It was found that the measured jitter varied approximately linearly with the expected jitter. The correlation coefficients between the expected and measured jitter were 0.97, 0.96, and 0.94 for the three MUAP ensembles studied.
机译:在线源模型下,模拟了同心针电机单元动作电位(MUAP)的集合。通过延迟正常随机变体延迟肌纤维作用电位(MFAPS)的启动来模拟神经肌肉结抖动。为了将MFAP贡献转换为对称,单峰值,正函数,逆平均电流滤波应用于集合的MUPAPS,产生复合重量函数。定义为在整体上的复合权重函数的标准偏差的抖动估计与用于延迟MFAP的启动的随机变量的标准偏差相关。由于正常抖动的顺序为25μsec,因此使用划分差插值增强了40μsec的基础采样间隔至大约2微秒。发现测量的抖动与预期的抖动大致线性变化。对于所研究的三个Muap合奏,预期和测量的抖动之间的相关系数为0.97,0.96和0.94。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号