A computer simulation study has been conducted to evaluate the effect of the interelectrode distance on the derivation of body surface. Laplacian maps from body surface potential distribution. A three-dimensional computer heart-torso model was used to simulate cardiac electrical activity, body surface potential map (BSPM) and body surface Laplacian map (BSLM). A two-site pacing protocol was used to generate two simultaneous myocardial events over the anterior epicardium. The derived body surface Laplacian was estimated from the potential difference between the potential at the center point and the averaged potential over the four surrounding points. The present simulation results suggest that BSLM may be estimated from the body surface potentials if the interelectrode distance is reasonably small.
机译:已经进行了计算机仿真研究以评估电极距离电极距离对体表衍生的影响。 Laplacian从体表电位分布的地图。三维计算机心脏躯干模型用于模拟心脏电活动,体表潜在地图(BSPM)和体表Laplacian地图(BSLM)。使用双地起搏方案在前心外表皮上产生两个同时心肌事件。衍生的体表Laplacian估计从中心点的电位与四个周围点上的平均电位之间的电位差估计。本仿真结果表明,如果电极互际距离相当小,则可以从体表电位估计BSLM。