首页> 外文会议>International Conference on Environmental Systems >Results From the Vehicle Cabin Atmosphere Monitor: A Miniature Gas Chromatograph/Mass Spectrometer for Trace Contamination Monitoring on the ISS and Orion
【24h】

Results From the Vehicle Cabin Atmosphere Monitor: A Miniature Gas Chromatograph/Mass Spectrometer for Trace Contamination Monitoring on the ISS and Orion

机译:车厢大气监测结果:微型气相色谱仪/质谱仪,用于在ISS和ORION上进行痕量污染监测

获取原文
获取外文期刊封面目录资料

摘要

Progress on the delivery of the Vehicle Cabin Atmosphere Monitor (VCAM) is reported. VCAM is an autonomous trace-species detector to be used aboard the International Space Station (ISS) for atmospheric analysis. The instrument is based on a low-mass, low-power miniature preconcentrator, gas chromatograph, and Paul ion trap mass spectrometer (PCGC/MS) capable of measuring volatile constituents in a space vehicle or planetary outpost at sub-ppm levels. VCAM detects and quantifies 40 target compounds at their 180-day Spacecraft Maximum Allowable Concentration (SMAC) levels. It is designed to operate autonomously, maintenance-free, with a self-contained carrier and calibration gas supplies sufficient for a one-year lifetime. Two flight units will be delivered for operation in the ISS EXPRESS rack.
机译:报道了车厢大气监测器(VCAM)交付的进展。 VCAM是一种自主痕量物种探测器,用于乘坐国际空间站(ISS)进行大气分析。该仪器基于低质量,低功耗微型预融合器,气相色谱仪和保罗离子阱质谱仪(PCGC / MS),其能够在亚PPM水平的空间车辆或行星前哨中测量挥发性成分。 VCAM以其180天的航天器最大允许浓度(SMAC)水平检测和量化40个目标化合物。它旨在自主,免维护,自包含的载体和校准气体供应,足以实现一生。两个飞行单位将在ISS Express Rack中运送操作。