A comparison of 2D and 3D edge detectors in semi automated measurements of chamber volumes using 3D echocardiographic laboratory phantom images

机译:使用3D超声心动图实验室幻影图像对腔室容积进行半自动测量的2D和3D边缘检测器的比较

获取原文

摘要

Chamber volume measurement is fundamental for assessing many indices of cardiac function. 3D echocardiography offers great potential for accurate measurement, but many automated or semi-automated techniques suffer from acoustic noise and artefact, which can be reduced by suitable choice of edge detectors. We compared the effects of 2D and 3D first order derivative edge detectors with similar size on accuracy and repeatability in six volume measurement methods of 3D echo images of a laboratory balloon phantom. Overall accuracy compared with known volumes was better for 3D (-1.0 ml; -4.0 ml to 2.0 ml 95% CI; P=0.53) than 2D (-11.6 ml; -14.8 ml to -8.5 ml 95% CI; P<0.001). Repeatability was better for 3D (SD=0.9 ml) than 2D (SD=1.5 ml) with difference in coefficient of variation of 0.6 (P<0.001). 3D operators show potential for better performance with 3D echo images.
机译:腔室容积测量对于评估许多心脏功能指标至关重要。三维超声心动图进行精确测量提供了巨大的潜力,但许多自动化或半自动化技术,从声学噪声和伪影,这可以通过边缘检测器的合适的选择可以降低苦。我们在实验室气球模型的3D回波图像的六种体积测量方法中,比较了大小相似的2D和3D一阶导数边缘检测器对准确性和可重复性的影响。与2D(-11.6 ml; -14.8 ml至-8.5 ml 95%CI)相比,3D(-1.0 ml; -4.0 ml至2.0 ml 95%CI;与已知体积相比,总体精度更好) )。 3D(SD = 0.9 ml)的重复性优于2D(SD = 1.5 ml),变异系数的差异为0.6(P <0.001)。 3D运营商显示出具有3D回波图像更好性能的潜力。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号