Validation of Pulse Transit Time Based Blood Pressure Estimation on Atrial Fibrillation Patients

机译:基于脉搏传播时间的心房颤动患者血压估计的验证

获取原文

摘要

Pulse transit time (PTT) based continuous cuff-less blood pressure (BP) monitoring has attracted wide interests owing to its potential in improving the control and early prevention for cardiovascular diseases. However, it is still impractical in large-scale clinical application due to the concern of BP measurement accuracy. Since such approach strongly relies on the PTT-BP model under certain theoretical assumptions, the accuracy would be affected by the vessel properties alterations induced by cardiovascular disorders. Atrial fibrillation (AF) is one of the most common cardiac diseases which often coexist with hypertension. The present study sought to examine the Impact of AF on the PTT and BP, validate the capability of PTT based cuff-less methods on AF patients. By investigating the PTT and BP on 74 critically ill patients with AF, we found that parameters including PTT, R-R interval and diastolic BP (DBP) were significantly changed when AF occurs, while the systolic BP (SBP) value and photoplethysmography intensity ratio (PIR) changed little. Further, by performing two cuff-less BP estimation method, we found that the estimated accuracy is decreased on PTT based method when AF occurs, but there is little change on PIR based method. The findings demonstrated that the impact of AF on PTT is significant, which would also influence the PTT-BP relationship. But the PIR would still be a predictive factor for BP estimation for AF patients.
机译:基于脉搏传播时间(PTT)的连续无袖带血压(BP)监测引起了广泛的关注,因为它具有改善心血管疾病的控制和早期预防的潜力。然而,由于对BP测量精度的关注,在大规模临床应用中仍然不切实际。由于这种方法在某些理论假设下强烈依赖PTT-BP模型,因此准确性将受到心血管疾病引起的血管特性改变的影响。心房颤动(AF)是最常见的心脏病之一,通常与高血压并存。本研究旨在探讨AF对PTT和BP的影响,验证基于PTT的无袖套方法对AF患者的能力。通过对74例重症房颤患者的PTT和BP进行调查,我们发现,发生AF时,包括PTT,RR间隔和舒张压(DBP)在内的参数发生了显着变化,而收缩压(SBP)值和光电容积描记强度比(PIR) )变化不大。此外,通过执行两个无袖带BP估计方法,我们发现,当发生AF时,基于PTT的方法的估计准确性降低,但是基于PIR的方法几乎没有变化。研究结果表明,AF对PTT的影响是显着的,这也将影响PTT-BP关系。但是,PIR仍将是AF患者进行BP评估的预测因素。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号