首页> 外文会议>Annual International Conference of the IEEE Engineering in Medicine and Biology Society >A clinical study of short-term sternal photoplethysmography: Recordings from patients with obstructive airways diseases
【24h】

A clinical study of short-term sternal photoplethysmography: Recordings from patients with obstructive airways diseases

机译:短期胸骨光学性分析的临床研究:阻塞性气道疾病患者的录音

获取原文

摘要

Traditionally, measurements of the oxygen saturation (SpO2) has been confound to the extremities. In this study, we therefore investigated the possibility for reliable estimation of clinically relevant SpO2 levels from photoplethysmography (PPG) obtained on the sternum of patients with obstructive airway diseases. We initiated the study with a calibration of a prototype sternal PPG sensor. In accordance with the ISO 80601-2-61:2011 guidelines, the calibration was conducted as a controlled desaturation study. We obtained a calibration accuracy of 1.75% which is well within the clinically and commercially accepted range. We then compared the SpO2 levels simultaneously obtained from the sternal PPGs and a commercially available finger pulse oximeter on 28 admitted patients with either asthma or Chronic Obstructive Pulmonary Disease (COPD). The Pearson correlation between the SpO2 levels estimated from the two body locations was found to be 0.89 (p<;0.05) and the mean system bias was only 0.052% with upper and lower limits of agreement of 2.5% and -2.4%, respectively. This finding is very promising for the future design of new sternum based patch technologies that might be able to provide continuous estimates of the SpO2 levels on critically or chronically ill patients.
机译:传统上,氧气饱和度(Spo2)的测量已经困扰到四肢。因此,在本研究中,我们研究了可靠地估计来自梗阻性气道疾病患者胸骨胸骨胸骨胸骨上获得的临床相关的SPO2水平的可能性。我们通过校准原型胸骨PPG传感器启动了该研究。按照ISO 80601-2-61:2011指导方针,校准被作为受控的去饱和研究进行。我们获得了1.75%的校准精度,在临床上和商业上接受的范围内。然后,我们将从胸骨PPG和哮喘或慢性阻塞性肺病(COPD)的患者的商业手指脉冲血氧计(COPD)进行比较从胸骨PPG和商业手指脉冲血氧计的孢子血管计。从两个身体位置估计的SPO2水平之间的Pearson相关性被发现为0.89(p <; 0.05),平均系统偏差仅为0.052%,分别为2.5%和-2.4%的上下限制。这一发现对于未来的新胸骨补丁技术设计非常有前途,可能能够在批判性或长期生病的患者上提供对SPO2水平的连续估计。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号