This study includes the design, electrical and physiological tests of an electrical stimulator that is able to excite nerves and muscles. The device consists of 3 different units which are stimulator, power supply and computer interface. The stimulator unit includes 3 parts which are; a voltage-current convertor by a power op amp, isolation and an external device trigger parts. The device has an external power supply unit in order to supply high voltage for power op amp. The system is controlled and driven by a computer interface designed in LabVIEW and the interface drives stimulator via audio output of a computer. Physiological tests were performed under the supervision of authorized people. Data of this study is compared with literature data.
机译:这项研究包括能够刺激神经和肌肉的电刺激器的设计,电学和生理学测试。该设备由3个不同的单元组成,分别是刺激器,电源和计算机接口。刺激器单元包括3个部分:电压-电流转换器由功率运算放大器,隔离器和外部设备触发部分组成。该设备具有一个外部电源单元,以便为功率运算放大器提供高电压。该系统由LabVIEW设计的计算机接口控制和驱动,该接口通过计算机的音频输出驱动刺激器。生理测试是在授权人员的监督下进行的。将本研究的数据与文献数据进行比较。