首页> 外文会议>International Conference on BioMedical Engineering and Informatics >Nonlinear relationship between systolic blood pressure and pulse transit time in anesthetized dogs
【24h】

Nonlinear relationship between systolic blood pressure and pulse transit time in anesthetized dogs

机译:麻醉犬收缩压与脉搏传播时间的非线性关系

获取原文

摘要

Pulse wave transit time (PTT), as measured between the ECG R-wave and the arterial pulse wave at a distal site, has been previously reported to linearly correlate with blood pressure (BP). This study reinvestigated the relationship where various doses of epinephrine were administered to three anesthetized dogs to invoke a wide range of hypotension. The invasive systolic blood pressure in the left ventricle, the noninvasive photoplethysmogram pulse wave and the ECG of lead II were recorded synchronously. The results indicated that the relationship between the systolic BP and the PTT was nonlinear and logarithmic: P=b-b lnT, where b and b are coefficients. The estimated BP using the nonlinear relationship was highly correlated with the measured BP; the correlation coefficients were often greater than 0.96 in the 36 trials. The root mean square of the estimation error was less than 6.53 mmHg. However, the coefficients b and b varied between trials, doses, and dogs. An analysis of the error contribution revealed that the estimation error increased linearly with the coefficient errors, Δb and Δb, and with the PTT error, ΔT. The definition of the pulse arrival time (PAT) also affected the performance of the estimation. The performance based on five PAT definitions was compared. This study shows that caution is necessary when using this relationship to monitor BP.
机译:先前已经报道了在远侧部位的ECG R波和动脉脉搏波之间测得的脉搏波传播时间(PTT)与血压(BP)呈线性相关。这项研究重新研究了对三只麻醉狗使用各种剂量肾上腺素以引起广泛的低血压的关系。同步记录左心室的有创收缩压,无创光体积描记脉搏波和铅II的心电图。结果表明,收缩压与PTT的关系是非线性和对数的:P = b-b lnT,其中b和b是系数。使用非线性关系估算的BP与测得的BP高度相关。在36个试验中,相关系数通常大于0.96。估计误差的均方根小于6.53 mmHg。但是,系数b和b在试验,剂量和狗之间有所不同。对误差贡献的分析表明,估计误差随着系数误差Δb和Δb以及PTT误差ΔT线性增加。脉冲到达时间(PAT)的定义也影响估计的性能。比较了基于五个PAT定义的性能。这项研究表明,使用这种关系来监测血压时必须谨慎。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号