The National Institute of Standards and Technology (NIST) uses a bi-directional 20 L-displacement piston prover as its primary standard for measuring hydrocarbon liquid flows ranging from 1.86 x 10~5 m3/s (0.3 gpm) to 2.6 x 10"3 m3/s (40 gpm). Our analysis that shows that the prover's uncertainty over this flow range is 0.074 % (it = 2, corresponding to a 95 % confidence interval). Using a dual rotor turbine meter as the transfer standard, we compare the new 20 L piston prover standard with NIST's other hydrocarbon liquid standards and NIST's water flow standard. The results are consistent with the presented uncertainty analysis.
机译:美国国家标准技术研究院(NIST)使用双向20 L排量活塞检定仪作为其主要标准,用于测量烃类液体流量,范围从1.86 x 10〜5 m3 / s(0.3 gpm)到2.6 x 10“ 3 m3 / s(40 gpm)。我们的分析表明,在该流量范围内,证明方的不确定度为0.074%(它等于2,对应于95%置信区间)。使用双转子涡轮流量计作为传输标准,我们进行了比较新的20 L活塞检验器标准品与NIST的其他碳氢化合物液体标准品和NIST的水流量标准品相符,结果与提出的不确定性分析相符。