We present a dual-band HF and UHF fully-integrated implantable neuroprosthetic testbed. This testbed includes a custom implanted device as well as an external system based on a commercially available USRP B210 software defined radio (SDR) platform. The implanted device integrates a BPSK backscatter uplink rate of 5 Mbps, an HF WPT efficiency of 47% with a power consumption of 1.332 milliwatt. The implanted device measures 25 mm diameter and has a total thickness of 2.8 mm including the printed circuit substrate, antenna, all circuitry, and silicone encapsulation. It supports up to 10 neural and 4 electromyogram (EMG) channels with a sampling rate of 26.1 kHz for the neural channels and 1.628 kHz for the EMG channels. The communication link is shown to have 0% packet error rate at an implant depth of up to 2.5 cm.
机译:我们提出了双带HF和UHF全综合的植入神经高原试验台。该测试平台包括自定义植入设备以及基于市售USRP B210软件定义的无线电(SDR)平台的外部系统。植入装置集成了5 Mbps的BPSK反向散射上行率,HF WPT效率为47 %,功耗为1.332毫瓦。植入装置直径为25毫米,总厚度为2.8mm,包括印刷电路基板,天线,所有电路和硅胶封装。它支持最多10个神经和4个电灰度(EMG)通道,采样率为26.1kHz的神经通道,1.628 kHz为EMG通道。通信链路显示在植入深度高达2.5厘米的植入深度下具有0 %的分组错误率。