however the resolution of gel dosimeter is strongly dependant to gel composition, The proper parameters and selection of optimum method in dose response readout can reduce noise and improve some important features such as contrast and spatial resolution considerably. One method for dose response readout in gel dosimetry is variation of magnetic properties of proton using magnetic resonance imaging (MRI). A major problem in this method, is the serious reduction of contrast in edges when results in lost of dose in those region. In this study a new method contrast enhancement of gel dosimeter with MR image is presented, in doses lower than 17 Gy with imaging in water medium, a small reduction in spatial resolution is exchange to considerable increase of contrast in R2 map is important.
机译:但是,凝胶剂量计的分辨率在很大程度上取决于凝胶的组成。在剂量响应读数中选择合适的参数和选择最佳方法可以降低噪音并显着改善一些重要功能,例如对比度和空间分辨率。凝胶剂量测定中剂量响应读数的一种方法是使用磁共振成像(MRI)改变质子的磁性。该方法的主要问题是当在那些区域中失去剂量时边缘的对比度严重降低。在这项研究中,提出了一种用MR图像增强凝胶剂量计对比度的新方法,在水介质中以低于17 Gy的剂量成像时,将空间分辨率的小幅降低交换为R2图中对比度的显着提高非常重要。