In vivo measurements of the dose delivered by kilovoltage cone-beam CT during prostate radiation therapy

机译:前列腺放射治疗期间千伏锥束CT的体内剂量测量

获取原文

摘要

The usefulness of CBCT is proven, but this technique can deliver a significant dose to the patient when it is often practiced. The dose is very dependent on the equipment and the parameters used. A study has been undertaken on an anthropomorphic phantom and on patients irradiated for prostate cancer, in order to evaluate the mean dose delivered to the prostate and rectum (organ at risk). Images have been taken using a Varian equipment (model VZW2940FH2-02 -detector Pax-Scan 4030 CB made of aSi) in "half bow-tie" mode, maximum field size, 125 kV, 80 mA, 25 ms. Dosimetry was performed using Gafchromic films and Harshaw - TLD 700H calibrated by comparison with a Keithley ion chamber type 90035.A good agreement has been found between films and TLD results. Films have shown that the dose distribution is very inhomogeneous and that the dose at the center of the phantom depends both on the explored zone thickness and on its position relative to the beam axis. The doses measured at the phantom surface (TLD and films) were maximum at the entrance of the anterior-posterior axis, and their arithmetic mean is equal, or a little greater than the doses measured at mid-thickness of the phantom and at the level of the rectum. After discussion with the medical team, it has been decided that the average between the anterior and posterior skin doses will be systematically evaluated in vivo, and taken as reference for correcting the dose delivered to the patient during radiation therapy. Study performed on 10 patients has shown mean doses per CBCT varying from 5.13 ± 0.23cGy to 7.28cGy ± 0.23cGy. Conclusion is that the doses delivered by CBCT are not negligible and should be taken into account. The RBE value to be applied for kilovoltage x-rays is questionable.
机译:CBCT的有用性已被证明,但是这种技术在经常实践时可以为患者提供可观的剂量。剂量非常取决于设备和所使用的参数。为了评估传递到前列腺和直肠(处于危险中的器官)的平均剂量,已经对拟人模型和接受前列腺癌照射的患者进行了研究。使用Varian设备(型号为VZW2940FH2-02 -detector Pax-Scan 4030 CB,由aSi制成)以“半领结”模式拍摄图像,最大场大小为125 kV,80 mA,25 ms。使用Gafchromic膜和Harshaw-TLD 700H(与Keithley离子室90035型进行比较)进行校准,进行剂量测定。 电影和TLD结果之间已经找到了很好的协议。胶片显示剂量分布非常不均匀,体模中心的剂量取决于探查区域的厚度及其相对于光束轴的位置。在幻影表面(TLD和胶片)处测得的剂量在前后轴的入口处最大,其算术平均值等于或略大于在幻影中等厚度和水平处测得的剂量直肠与医疗团队讨论后,已决定将在体内系统评估前后皮肤剂量之间的平均值,并将其作为校正放射治疗期间输送给患者的剂量的参考。对10位患者进行的研究显示,每个CBCT的平均剂量从5.13±0.23cGy到7.28cGy±0.23cGy不等。结论是CBCT的剂量不可忽略,应予以考虑。用于千伏X射线的RBE值值得怀疑。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号