首页> 外文会议> >Forward-looking ultrasound imager using convex-shaped 1-3composite transducer for intravascular treatment
【24h】

Forward-looking ultrasound imager using convex-shaped 1-3composite transducer for intravascular treatment

机译:使用凸形1-3复合换能器进行血管内治疗的前瞻性超声成像仪

获取原文

摘要

An intravascular forward-looking ultrasound probe (FLUP) of 3 mm in diameter has been developed. We used the FLUP as acoustic sources and receivers to make an ultrasound forward-looking image, but many artifacts were observed because of strong directivity of the FLUP. To confirm the affection of directivity in ultrasound forward-looking images, laser induced breakdown (LIB) was used as a low directional transmitter and a FLUP was used as a receiver, a fine image was successfully obtained because the LIB can provide an almost, completely spherical ultrasound pulse wave. For decreasing the directivity of FLUP, a convex-shaped 1-3composite PZT transducer (0.7 mm in diameter and 0.75 mm curvature radius) was fabricated, and low directivity feature was confirmed. A probe which has eight separated 1-3composite convex-shaped transducers were fabricated. The probe is 5 mm in diameter and a working channel is 0.5 mm in diameter. Acoustic pressure was measured using a hydrophone which was located 10 mm in front of the FLUP. The emitting acoustic pressure of 40 Pa was realized with the fabricated probe. When the probe was used as acoustic sources and receivers, echo signal of an aluminum plate was observed.
机译:已开发出直径3 mm的血管内前瞻性超声探头(FLUP)。我们使用FLUP作为声源和接收器来制作超声前视图像,但是由于FLUP具有很强的指向性,因此观察到了许多伪像。为了确认方向性在超声前视图像中的影响,激光诱导击穿(LIB)被用作低方向发射器,FLUP被用作接收器,由于LIB可以提供几乎,完全的图像,因此成功获得了精细的图像。球形超声脉搏波。为了降低FLUP的方向性,制造了凸形的1-3复合PZT换能器(直径0.7 mm,曲率半径0.75 mm),并确认了低方向性特征。制作了具有八个分离的1-3个复合凸形换能器的探头。探头的直径为5毫米,工作通道的直径为0.5毫米。使用位于FLUP前方10毫米的水听器测量声压。用所制造的探针实现了40Pa的发射声压。当探头用作声源和接收器时,观察到铝板的回波信号。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号