首页> 外文会议> >Neuromonitoring using compound action potential and cochlear blood flow during cochlear hypoxia for neurosurgery
【24h】

Neuromonitoring using compound action potential and cochlear blood flow during cochlear hypoxia for neurosurgery

机译:在神经外科耳蜗缺氧期间使用复合动作电位和耳蜗血流量进行神经监测

获取原文

摘要

A neuromonitoring system using compound action potential (CAP) and cochlear blood flow (CBF) is developed to assess the functional status of the cochlea during hypoxia. The system uses a specially designed optic probe capable of acquiring CAP and CBF signals and modules to process and display these signals. The system is tested in rabbits by creating cochlear hypoxia by manual compression of the auditory nerve and acoustic artery. Results show that CAP and CBF measurements can provide almost real time information on the status of the cochlea under compromising situations.
机译:开发了使用复合动作电位(CAP)和耳蜗血流(CBF)的神经监测系统,以评估缺氧期间耳蜗的功能状态。该系统使用专门设计的光学探头,该探头能够采集CAP和CBF信号,并使用模块来处理和显示这些信号。通过人工压缩听觉神经和听觉动脉产生耳蜗缺氧,在兔子中对该系统进行了测试。结果表明,CAP和CBF测量可以在折衷情况下提供几乎实时的耳蜗状态信息。