Radiation dose during CT scan with PET/CT clinical protocols

机译:使用PET / CT临床规程进行CT扫描期间的辐射剂量

获取原文

摘要

The purpose of this study was to evaluate the radiation doses during CT scans by changing tube voltage and tube current, and to estimate the radiation dose for our clinical protocols during transmission measurement in PET/CT scans. Radiation doses were evaluated for Philips GEMINI 16 slices PET/CT and GE DSTe 8 slices PET/CT systems. A CTDI was measured with standard head and body CT dosimetry phantoms in a variety of CT tube voltage and tube current. A pencil ionization chamber with an active length of 100 mm, and electrometer were used for the CTDI measurement. The measurement is carried out at the free-in-air, center and periphery. The average dose was calculated by the weighted CTDI (CTDIw = 1/3CTDI100,c + 2/3CTDI100,p). Specific organ doses were measured with our clinical protocols for transmission measurement with the Philips PET/CT system and GE PET/CT system with TLDs and anthropomorphic phantom. The TLDs used for measurements were selected for an accuracy of plusmn5% and calibrated in an X-ray radiation field. The organ or tissue was selected by the recommendations of ICRP 60. The radiation dose measured by using standard CTDI phantoms during a CT scan is affected by the tube voltage and the tube current. The measured effective doses with transmission measurements for the Philips system were ranged from 0.14 mSv to 55.54 mSv. The measured effective doses with transmission measurements for GE system were ranged from 0.27 mSv to 24.83 mSv. Radiation dose for clinical protocols during transmission measurements in the PET/CT system can be measured using CTDI phantom with ionization chamber and anthropomorphic phantom with TLDs. Further studies need to be performed to find optimal CT acquisition protocols for reducing the patient exposure with same image quality.
机译:这项研究的目的是通过改变管电压和管电流来评估CT扫描过程中的辐射剂量,并评估我们在PET / CT扫描中进行透射测量期间的临床方案的辐射剂量。对飞利浦GEMINI 16片PET / CT和GE DSTe 8片PET / CT系统的辐射剂量进行了评估。在各种CT管电压和管电流中,使用标准的头部和身体CT剂量学模型来测量CTDI。有效长度为100 mm的笔电离室和静电计用于CTDI测量。测量是在空气中,中心和外围进行的。通过加权CTDI(CTDI w = 1 / 3CTDI 100,c + 2 / 3CTDI 100,p )计算平均剂量。根据我们的临床方案测量特定器官的剂量,并使用飞利浦PET / CT系统和带有TLD和拟人模型的GE PET / CT系统进行透射测量。选择用于测量的TLD的精度为±5%,并在X射线辐射场中进行校准。器官或组织是根据ICRP 60的建议进行选择的。在CT扫描过程中使用标准CTDI体模测量的辐射剂量受电子管电压和电子管电流的影响。对于飞利浦系统,通过透射测量所测得的有效剂量为0.14 mSv至55.54 mSv。通过GE系统的透射测量所测得的有效剂量为0.27mSv至24.83mSv。可以使用带电离室的CTDI体模和带TLD的拟人体模来测量PET / CT系统中传输测量过程中临床方案的辐射剂量。需要进行进一步的研究以找到最佳的CT采集方案,以减少具有相同图像质量的患者暴露。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号