首页> 外文会议> >Validation and evaluation of model-based cross-talk compensation method in simultaneous /sup 99m/Tc stress and /sup 201/Tl rest myocardial perfusion SPECT
【24h】

Validation and evaluation of model-based cross-talk compensation method in simultaneous /sup 99m/Tc stress and /sup 201/Tl rest myocardial perfusion SPECT

机译:/ sup 99m / Tc应力和/ sup 201 / Tl静息心肌灌注SPECT的基于模型的串扰补偿方法的验证和评估

获取原文

摘要

Simultaneous acquisition of /sup 99m/Tc stress and /sup 201/Tl rest myocardial perfusion SPECT has several potential advantages, but the image quality is degraded by cross-talk between the Tc and Tl data. We have previously developed down-scatter and Pb X-ray models for cross-talk compensation. In this work, we validated these models by comparing the crosstalk from /sup 99/Tc to the Tl window calculated using our models with the cross-talk simulated using a combination of the SimSET-MCNP Monte Carlo simulation codes. We have also evaluated the model-based cross-talk compensation method using both simulated data from MCAT phantoms and experimental data from a physical phantom with a myocardial defect. In the studies the Tl distributions were reconstructed from cross-talk contaminated data without cross-talk compensation, with compensation using the model-calculated cross-talk or using the true cross-talk, and were compared with that from uncontaminated Tl data. Results show that the models gave good estimates of both the down-scatter photons and Pb X-rays in the simultaneous dual isotopes myocardial perfusion SPECT. Also, the results show that model-based cross-talk compensation method provided image quality significantly improved as compared to no compensation and very close to that from the separate acquisition.
机译:同时获取/ sup 99m / Tc应力和/ sup 201 / Tl静息心肌灌注SPECT具有多个潜在优势,但是图像质量因Tc和Tl数据之间的串扰而降低。我们以前已经开发了向下散射和Pb X射线模型以进行串扰补偿。在这项工作中,我们通过将来自/ sup 99 / Tc的串扰与使用我们的模型计算出的Tl窗口与使用SimSET-MCNP蒙特卡洛模拟代码的组合进行模拟的串扰进行比较,从而验证了这些模型。我们还评估了基于模型的串扰补偿方法,该方法同时使用了来自MCAT体模的模拟数据和来自具有心肌缺陷的物理体模的实验数据。在研究中,从没有串扰补偿的串扰污染数据重建T1分布,使用模型计算的串扰或使用真实串扰进行补偿,并将其与未污染的T1数据进行比较。结果表明,该模型对同时存在双同位素心肌灌注SPECT的向下散射光子和Pb X射线给出了良好的估计。而且,结果表明,与没有补偿相比,基于模型的串扰补偿方法可显着提高图像质量,并且与单独采集的图像质量非常接近。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号