首页> 外文会议> >A biomechanical study of a cervical spine stabilization device
【24h】

A biomechanical study of a cervical spine stabilization device

机译:颈椎稳定器的生物力学研究

获取原文

摘要

The biomechanical properties of surgically destabilized cervical spine motion segments are evaluated. The segments were fixed with spinal fixation plates and tested. The Roy-Camille spinal fixation plates were utilized as the stabilizing device. Results are compared with intact motion segments. It is shown that segments C3-4 and C4-5, when severely destabilized by a surgically created defect, can effectively be restabilized through proper application of Roy-Camille spinal fixation plates. In flexion/extension and torsion the passive range of motion is virtually eliminated in the plated motion segments. Tests to failure of these severely destabilized, plated spines demonstrate failure by screw pull-out.
机译:评估了手术不稳定的颈椎运动节段的生物力学特性。用脊柱固定板固定节段并进行测试。使用Roy-Camille脊柱固定板作为稳定装置。将结果与完整的运动段进行比较。结果表明,当C3-4段和C4-5段由于手术造成的缺损而严重不稳定时,可以通过正确使用Roy-Camille脊柱固定板来有效地重新稳定。在弯曲/伸展和扭转中,在电镀的运动节段中实际上消除了被动的运动范围。对这些严重不稳定的镀层刺的失效进行的测试表明,由于螺钉拔出而导致的失效。