Summary form only given. A special data acquisition unit has been developed which allows to process on-line not only the mean-phasic Doppler shift signal but also to analyze the spectral contents and the received A-mode signal. The probe consists of a 3f-catheter which is activated by a pulsed transmitter-receiver (Millar). The carrier frequency is 20 MHz and the impulse repetition rate is 62.5 kHz. Two modular A/D converters are used. One digitizes directly the Doppler shift frequencies in the range less than 20 kHz (sampling rate 100 kHz, resolution 12 bit); the other digitizes the RF-signal (sampling rate 200 MHz, resolution 8 bit). For further processing, data are transferred via separate bus systems to an IBM/AT compatible personal computer. An in vitro model has been developed to test the dependence of received signals on stenoses and different flow vehicles.
机译:仅提供摘要表格。已经开发了一种特殊的数据采集单元,该单元不仅可以在线处理均相多普勒频移信号,而且还可以分析频谱内容和接收到的A模式信号。探头由3f导管组成,该导管由脉冲发射器-接收器(Millar)激活。载波频率为20 MHz,脉冲重复频率为62.5 kHz。使用了两个模块化A / D转换器。一个人直接将小于20 kHz范围内的多普勒频移数字化(采样率100 kHz,分辨率12位);另一个数字化RF信号(采样率200 MHz,分辨率8位)。为了进行进一步处理,数据通过单独的总线系统传输到IBM / AT兼容的个人计算机。已经开发了体外模型来测试接收信号对狭窄和不同流动媒介的依赖性。