Our paper summarizes experiments for measuring the accuracy of deformable 2D-3D registration between sets of simulated x-ray images (DRR's) and a statistical shape model of the pelvis bones, which includes x-ray attenuation information ("density"). In many surgical scenarios, the images contain a truncated view of the pelvis anatomy. Our work specifically addresses this problem by examining different selections of truncated views as target images. Our atlas is derived by applying principal component analysis to a population of up to 110 instance shapes. The experiments measure the registration error with a large and truncated FOV. A typical accuracy of about 2 mm is achieved in the 2D-3D registration, compared with about 1.4 mm of an "optimal" 3D-3D registration.
机译:我们的论文总结了用于测量一组模拟X射线图像(DRR)与骨盆骨骼的统计形状模型之间的可变形2D-3D配准的精度的实验,其中包括X射线衰减信息(“密度”)。在许多外科手术场景中,图像包含骨盆解剖的截断视图。我们的工作通过检查截断视图的不同选择作为目标图像来专门解决此问题。我们的图集是通过对多达110个实例形状的总体应用主成分分析而得出的。实验以较大且截断的FOV测量配准误差。 2D-3D配准的典型精度约为2 mm,而“最佳” 3D-3D配准的精度约为1.4 mm。